After financing, capacity expansion and listing preparation, what will drive the next round of growth of domestic CT tubes?
In 2026, the CT X-ray tube industry is low-profile yet bustling.
It is low-profile because CT X-ray tubes do not carry the futuristic sci-fi narrative that comes with sectors like brain-computer interface and AI for Science. Hidden inside the complete CT scanner, it silently withstands the harsh tests brought by high temperature, high pressure and high-speed rotation. Its progress is more reflected in details such as materials, bearings, manufacturing processes and product reliability — these engineering breakthroughs that are not easy to externalize and difficult to be perceived by the public are strong support for the stable operation of CT equipment.
It is bustling because the actions of capital, regulation and the global supply chain this year are intensively falling on this X-ray tube —
In October, Zhishu Technology announced the completion of financing of over 200 million yuan and fully entered the stage of listing preparation;
In October, Neusoft Medical and Chaoqun Testing officially launched the DSA X-ray tube project for the foundation of Beijing's high-precision and advanced industries, carrying out core technology research around DSA X-ray tubes to promote the localization of core components of high-end medical imaging equipment;
In July, the equity change of Chaoqun Testing's Series D+ round was finalized, with support from industrial capital and national-level funds;
In April, Haogerui completed financing of tens of millions of yuan;
In January, Weirui Technology completed financing of nearly 100 million yuan;
In January, Yiteng Technology completed nearly 100 million yuan in Series A financing.
Arterial Network observed that "liquid metal bearings" have become the key development direction of the above-mentioned enterprises. At the same time, the focus of capital use is also shifting backward: from early-stage R&D and factory construction, it has gradually turned to automatic production line construction, capacity expansion, product pipeline extension and overseas market layout. This means that domestic CT X-ray tube enterprises are moving from the stage of technological breakthrough and product verification to large-scale production and commercial expansion.
The external environment of the CT X-ray tube industry is also changing. On October 3, the Ministry of Commerce of China suspended the anti-dumping investigation on imported related medical CT X-ray tubes originating in the United States and India.
Earlier, on August 10, Teledyne, a global industrial technology enterprise, announced that it planned to acquire Varex Imaging, a global supplier of core X-ray components, for about 1.1 billion US dollars. Varex has long provided core components such as X-ray tubes and flat-panel detectors for global medical imaging equipment manufacturers. Under the background that the global imaging core components are accelerating to concentrate on leading enterprises, this change will also bring new variables to the supply chain layout of Chinese CT complete machine manufacturers.
In terms of product supervision, the Center for Medical Device Evaluation of the National Medical Products Administration released the Guiding Principles for Registration Review of Reliability Evaluation of Tube Assemblies for X-ray Computed Tomography Equipment on February 28, further clarifying the preparation and evaluation requirements for reliability research data of CT tube assemblies. For core components that directly affect the imaging performance and use stability of CT equipment, reliability evaluation is becoming an important part of product registration review.
Under the superposition of a series of events, the CT X-ray tube industry has released several noteworthy signals: domestic capital continues to increase investment, and enterprises begin to accelerate large-scale layout; product reliability evaluation requirements are further clarified; the global imaging core component industry is also undergoing new integration.
When domestic CT X-ray tube enterprises gradually cross the thresholds of technology R&D and product registration, what key capabilities will the next stage of competition focus on? Will the long service life brought by liquid metal bearings become a "tightening spell" restricting the sales growth of X-ray tube enterprises? After liquid metal bearings become the common choice, where are the gaps between different enterprises? How can enterprises maintain profits after prices continue to fall...
Focusing on these issues, Arterial Network interviewed 5 X-ray tube enterprises that obtained financing this year and their investors, in order to look forward to the judgment of the next stage of industry development.
Why are all players focusing on liquid metal bearings? The answer lies in the laws of physics
To understand why domestic X-ray tube enterprises collectively bet on liquid metal bearings, we need to return to the working principle of the X-ray tube itself.
The core function of the X-ray tube is to generate X-rays: electrons emitted by the filament bombard the anode target disk under a high voltage of tens of thousands to hundreds of thousands of volts, only about 1% of the energy is converted into X-rays, and the remaining 99% is turned into heat. In order to prevent the target disk from being melted through instantly, the anode must rotate at a high speed to keep the landing point of the electron beam moving — this is the basic logic of the rotating anode X-ray tube.
The higher the number of CT rows, the higher the requirements for ray output and gantry rotation speed. For every 1% increase in the gantry speed, the centrifugal force borne by the entire X-ray tube increases accordingly; at the same time, the anode target disk itself rotates at a high speed under high temperature. The rolling contact of ball bearings is essentially point stress. When the load and speed exceed the material limit, the ball bearing cannot perform its function; the liquid metal bearing uses a liquid metal film to carry the rotating parts, with a large contact surface, evenly distributed load, and also takes into account thermal conduction.
Therefore, under the high-speed, high-power high-end CT working conditions, liquid metal bearings are the most mainstream and mature solution at present — this is the reason why high-end models with more than 64 rows are increasingly using X-ray tubes with liquid metal bearings in recent years.
But the liquid metal bearing itself is not a new technology. In 1989, Dunlee under Philips launched the first MRC X-ray tube with liquid metal spiral groove bearing, which was first used in its cardiovascular angiography system and entered the CT field about two years later. In the following two decades, GE did not complete the mass production and installation of self-developed liquid metal bearing X-ray tubes until the early 2010s. Because the threshold is so high, for a long time, the liquid metal bearing X-ray tubes used in domestic CT scanners with more than 64 rows almost all rely on imports.
However, it should be noted that the rise of liquid metal bearings does not mean the exit of ball bearings. Liquid metal bearings solve the problems that ball bearings cannot handle under high-end working conditions, rather than announcing the end of ball bearings. Ball bearings are still the main force in the current stock market.
The longer the service life of the X-ray tube, the lower the income of manufacturers?
Liquid metal bearings bring longer service life of X-ray tubes, which reduces costs for hospitals, but for X-ray tube manufacturers that mainly rely on one-time sales, it may mean that the replacement frequency decreases, which in turn affects revenue. So will the extension of X-ray tube service life really lead to the reduction of manufacturers' revenue?
In the view of the interviewees, this problem cannot simply deduce the change of enterprise income from the bearing type and product life.
First, from the technical facts, the service life has not been suddenly extended.
In the view of Yan Xun, founder of Haogerui, this problem is "very misleading". First, liquid metal bearings are not a new concept. Philips commercialized it in 1989, and GE applied it in 2015 — a large number of stocks in the market already use this technology, and there is no cliff where the service life suddenly changes from ball bearings to liquid metal; second, the core problem it solves is not only to extend the service life, but to meet the working condition requirements of higher gantry rotation speed and higher power, and make up for the capability boundary of ball bearings; third, from the empirical point of view, the service life of liquid metal bearing X-ray tubes used by international giants for models with more than 64 rows is not longer than that of ball bearing X-ray tubes used on 16-row models. Different models have different scanning loads and performance requirements, and the actual service life cannot be inferred only by the bearing type.
Then look at the pricing logic of hospitals: the market does not pay by "unit", but pays by service life.
Tang Zhihong, chairman of Chaoqun Testing, took a classic X-ray tube widely used in domestic 16-row and 32-row CT scanners as an example: the service life of some imitation tubes is only 100,000 to 200,000 scan seconds, and the original Philips tube can reach two to three times of that, "but hospitals do not think Philips' products are expensive". For this reason, low-end ball X-ray tubes are now sold for 70,000 to 120,000 yuan, while liquid metal bearing X-ray tubes can be sold for 300,000 to 400,000 yuan — the longer the service life, the higher the unit price, and the better the gross profit.
Zhishu Technology converted this logic into "cost per scan second": for an X-ray tube with doubled service life and only 30% higher price, the unit cost is directly reduced by 35%; while a cheap X-ray tube that breaks down in half a year and is accompanied by shutdown losses "saves the procurement budget, but loses the inspection income and emergency order. The last thing hospitals fear is not the expensive X-ray tube, but the shutdown of CT scanners."
For enterprises themselves, the current competition is far from the stage of "worrying about too long service life".
Nie Xipeng, founder of Weirui Technology, said frankly that the domestic CT X-ray tube industry as a whole is still in the stage of improving product performance and reliability, and there is still a distance from the concern that "too long service life will affect replacement sales". "In terms of trend, the improvement of product life will indeed lead to the decline of replacement frequency, but whoever can make products better first, make hospitals and customers more assured and operate more efficiently, will have more opportunities to obtain continuous orders and market recognition." He believes that the market share concentration brought by the advantages of service life and reliability is enough to hedge the impact of the decline in replacement frequency of single equipment.
If we extend the time line, the growth point should not only focus on "tube replacement".
Zou Yun, founder of Yiteng Technology, believes that extending the service life of X-ray tubes through technological progress and manufacturing capability improvement is the development direction that the industry should adhere to. In the short term, the extension of X-ray tube service life may reduce the replacement frequency of stock CT equipment. But for hospitals, one less tube replacement means one less shutdown caused by tube replacement. This is exactly the value that X-ray tube enterprises should provide for hospitals.
More importantly, new clinical demands are also promoting the development of X-ray tube technology. For example, in order to reduce image blur caused by heart movement, some CT gantries only need about 0.25 seconds to rotate one circle, which puts forward higher requirements for the mechanical bearing capacity of X-ray tubes. Perfusion, multi-phase enhancement and other examinations require multiple exposures, which also puts forward higher requirements for the output stability and heat dissipation capacity of X-ray tubes. For X-ray tube enterprises, developing new technologies around these clinical demands is the growth opportunity in the next stage.
The calculation of market capacity is not difficult. Lu Xiaobo, founder of Hongchuang Capital, estimates that the domestic CT inventory has reached tens of thousands of units. Even if estimated according to the two-year replacement cycle, the annual replacement demand is tens of thousands of units; and the unit price of liquid metal bearing X-ray tubes is much higher than that of ball X-ray tubes, and the same number of replacement units corresponds to several times the market value. Coupled with the loss caused by high-intensity use in large hospitals and thousands of new installations every year, "the replacement capacity of the X-ray tube market is very high every year, and the domestic market is a 100-billion-level or even larger market".
"From a social perspective, long service life reduces the cost of national medical equipment; from an enterprise perspective, after the market expands, you can also make money by relying on the efficiency of production, operation and management. Just like the semiconductor industry, after breaking through technical problems at first, it finally has to move to operational efficiency." Tang Zhihong concluded.
Convergent technical route ≠ convergent capability, the real barrier lies in engineering and reliability
The technical route of liquid metal bearings can converge, but whether it can be produced stably and in large quantities and withstand tests in actual use is the capability that domestic X-ray tube enterprises need to prove in the next stage.
Tang Zhihong, chairman of Chaoqun Testing, mentioned in the interview that the real difficulty of liquid metal bearings is not only designing the bearing, but also the subsequent filling process, manufacturing process and whole tube integration. The design of the bearing will affect the operation life, the filling process is related to the yield and long-term use performance, and the details in the manufacturing process will also affect the final product quality.
"Liquid metal bearing is not an isolated component. Even if the enterprise has mastered the bearing technology, it needs to combine it with the whole tube design, material selection, processing manufacturing and assembly process to form a truly deliverable product. This is also the special feature of the CT X-ray tube industry. It is not an industry where a single technological breakthrough can solve all problems." Tang Zhihong said.
Taking the anode target disk as an example, its material selection, structural design and welding quality will affect the heat capacity, thermal shock resistance and long-term use performance; the consistency of cathode emission characteristics is related to the stability of the focus and imaging quality; vacuum packaging, high-voltage insulation, heat dissipation management and other links will also jointly affect the operation state of the X-ray tube. Even the common ignition failure may involve multiple factors such as raw materials, structural design, manufacturing process and high-voltage management.
Therefore, the real technical barrier of X-ray tubes is the system engineering capability that integrates materials, machinery, vacuum, electronics, thermal management and manufacturing processes. This also explains why making a qualified sample tube in the laboratory is completely different from producing hundreds or even thousands of stable and reliable products in batches.
Hu Yinfu, founder of Zhishu Technology, also expressed a similar view: "In addition to bearings and several key technical links, batch consistency control and reliability data analysis of products are also crucial. Any of the three variables of part processing precision, cleanliness and surface finish drifts in the batch, and the yield will collapse. The laboratory index is "the best one", and the mass production index is "the deviation between the 1000th unit and the first unit". This level does not compete for R&D capability, but for the manufacturing system."
Yan Xun, founder of Haogerui, summed up the gap between enterprises as two important indicators: mass production yield and product reliability. The former determines whether an enterprise can transform R&D achievements into sustainably deliverable commodities, and the latter determines whether products can operate stably for a long time in the real hospital use environment.
In his view, the industry is undergoing changes from samples to commodities, from parameter substitution to clinical substitution, and from smooth installation to long-term stable service. Behind these three changes, it points to the same question: whether domestic X-ray tubes can truly meet the clinical use needs of hospitals.
For hospitals, purchasing X-ray tubes is not simply comparing a set of paper parameters. Even if two X-ray tubes are marked with the same heat capacity, it does not mean that they have exactly the same continuous scanning capability. Factors such as cooling system design, tube core structure, and focus characteristics will affect the actual use performance. More critically, hospitals need X-ray tubes that can stably output X-rays, meet daily scanning needs, and ensure image quality. For hospitals with a large number of patients, frequent failures and equipment shutdown not only mean replacement costs, but also affect inspection arrangements and operational efficiency.
"Therefore, the competition logic of X-ray tube products is different from that of ordinary parts. Performance is the premise of entering the market, and price advantage is meaningful only when it is built on reliability". Yan Xun summed it up as "performance comes first, price comes second": product performance needs to reach the level that is not inferior to imported brands and original products. On this basis, price advantage can be truly transformed into competitiveness. Therefore, domestic X-ray tube enterprises can not only prove that they "can make it", but also use real installation performance, failure rate, service life and batch delivery data to prove that they "can be used reliably".
The Guiding Principles for Registration Review of Reliability Evaluation of Tube Assemblies for X-ray Computed Tomography Equipment released in February 2026 also provides a clearer evaluation basis for the industry. It puts forward technical requirements around reliability research and verification, and promotes enterprises to carry out failure mode analysis, life evaluation and reliability tests more systematically.
The core changes of the new guidelines include: adopting the "three-element" reliability index to replace the original single expected service life — within the service period of Z years, under the condition of confidence Y, the reliability reaches X, the three figures are indispensable and must be written into the manual; providing two evaluation paths of preset index verification and life distribution model backstepping, both of which allow accelerated tests; the sample size gives a clear calculation method and is no less than 5; and requires the use of FMECA and other methods to classify and manage failure modes such as filament short circuit, target disk cracking, rotating anode jamming, and ignition.
After these requirements are truly implemented, "reliability" will no longer be a qualitative description at the publicity level, but a verifiable and checkable technical item. From this perspective, the competition of domestic CT X-ray tubes is essentially a comprehensive competition of engineering capability, manufacturing system and reliability data accumulation. Whoever can take the lead in establishing stable mass production capability and prove reliability with